CPSC official recall · 03506
CPSC, Sunrise Medical Announce Recall of 3-Wheel Hand Cycle
重要:以下安全事实保留 CPSC 官方英文原文,避免翻译改变含义。请通过右侧官方公告确认行动要求。
危险说明 Hazard
- There is a small risk that the quick release axles of the rear wheels do not lock correctly. If the quick release axle is not fully and properly engaged, the wheel could become detached while moving. A detached wheel could result in loss of control of the cycle and potential for serious injury or death, due to a fall or collision with another object.
受影响产品
3-Wheel Hand Cycle
Units:About 70
官方描述 Description
The recalled hand cycles come in a variety of colors and have "SOPUR" written on both sides of the bodywork.
事故与伤害 Incidents / Injuries
- Sunrise Medical has received one report, in which the quick release failed and the rider suffered minor injuries.
官方补救措施 Remedy
- Consumers should stop riding these cycles immediately and contact their local Sunrise Medical dealer for a free repair.
消费者联系信息
For more information, contact Sunrise Medical in Germany at 0049-7253-980-340 or by e-mail s.gerster@sunrisemedical.de -- Consumers also can visit the firm's web site at www.sunrisemedical.de
相关企业
制造商:Sunrise Medical of Heidelberg, Germany
制造国家/地区:Germany
零售商:Sunrise Medical dealers nationwide sold these hand cycles, from August 2001 to July 2002, for $3,300.