CPSC, Vernier Software & Technology Announce Recall of Go!™ Temp Temperature Sensors
危险说明 Hazard
- When connected to a computer, a static discharge to the sensor can cause the sensor to draw too much current from the computer, causing the sensor to get hot. The sensor can get hot enough to cause a burn to the skin or damage objects in contact with it.
受影响产品
官方描述 Description
Vernier's Go!™ Temp is a temperature sensor that connects directly to a computer USB port and collects data (temperature) from the probe. The sensors are used in science classrooms. They have serial numbers 0104XXXXXXXX through 2004XXXXXXXX. The serial number is found by connecting the sensor to the computer after installing Logger Lite software, which is included with the sensor. To determine the serial number, connect the sensor, start Logger Lite, pull down the Experiment menu, choose Set Up Sensors, and choose Go!Temp. The serial number will appear in the dialog box displayed on the computer screen.
事故与伤害 Incidents / Injuries
- Vernier reported one incident in which a teacher suffered burns to her palm.
官方补救措施 Remedy
- Consumers should stop using the Vernier Go!™ Temp included in this recall immediately and contact Vernier Software & Technology for a free replacement.
消费者联系信息
For additional information, contact Vernier Software & Technology at (888) 837-6437 between 7 a.m. and 6 p.m. PT Monday through Friday and on the firm's Web site at www.vernier.com/recall