CPSC official recall · 08603
GIA Instruments Recalls Polariscopes Due to Shock Hazard
重要:以下安全事实保留 CPSC 官方英文原文,避免翻译改变含义。请通过右侧官方公告确认行动要求。
危险说明 Hazard
- The polariscopes can have a ground wire missing, posing a risk of electrical shock to the user.
受影响产品
Illuminator Polariscopes 415000 and 415
Units:About 8,000
官方描述 Description
The recalled polariscopes are white in color with model number 415000 or 415 printed on the underside of the instrument. A polariscope is an instrument used by gemologists to assist in gem identification.
事故与伤害 Incidents / Injuries
- GIA has received four reports of consumers experiencing minor electric shock.
官方补救措施 Remedy
- Consumers should immediately stop using the recalled product and contact GIA Instruments to schedule a free retrofit of the Polariscope with a ground wire. GIA is notifying all consumers directly.
消费者联系信息
For additional information, contact GIA Instruments at (800) 421-8161 between 8 a.m. and 4 p.m. PT Monday through Friday or visit the firm's Web site at www.gia.edu
相关企业
制造商:GIA Instruments, of Carlsbad, Calif.
制造国家/地区:United States
零售商:GIA schools, GIA's Web site, product catalogs, distributors and trade shows nationwide from January 1996 through June 2007 for between $255 and $340.