CPSC official recall · 11067

Extech Instruments Recalls Digital Clamp and Multimeters due to Electrocution Hazard

召回日期:2010-12-14官方最后发布:2016-04-26适用地区:美国
重要:以下安全事实保留 CPSC 官方英文原文,避免翻译改变含义。请通过右侧官方公告确认行动要求。

危险说明 Hazard

  • When the battery runs low, the meters can fail to give an accurate voltage reading, resulting in the operator falsely believing the electrical power is off or low. This poses an electrocution hazard.

受影响产品

Digital clamp meters and multimeters
Units:5,100 units

官方描述 Description

This recall involves Extech digital clamp meters with model numbers EX612, EX613, EX622 and EX623 and multimeters with model numbers EX540, EX542 and EX570. "Extech" and the model number can be found on the front of the product and the serial number on the rear. Serial numbers that begin with "A" are not included in the recall.

事故与伤害 Incidents / Injuries

  • Extech received one report of a meter displaying an incorrect voltage reading. No injuries have been reported.

官方补救措施 Remedy

  • Consumers should immediately stop using the recalled digital clamp meters and multimeters and contact Extech for a free replacement meter.

Options:Replace

消费者联系信息

For additional information, contact Extech toll-free at (855) 239-8324 between 8:30 a.m. and 5 p.m. ET Monday through Friday or visit the company's website at http://www.extech.com/recall

相关企业

制造商:Extech Instruments, of Waltham, Mass.

制造国家/地区:China

零售商:Industrial and electrical distributors and wholesalers nationwide and at online tool and test equipment retailers from January 2008 to November 2010 for between $150 and $300.